An Investigation Into the Differences in Bone Density and Body Composition Measurements Between 2 GE Lunar Densitometers and Their Comparison to a 4-Component Model

An Investigation Into the Differences in Bone Density and Body Composition Measurements Between 2 GE Lunar Densitometers and Their Comparison to a 4-Component Model

Objective:

To assess the precision of the GE Lunar iDXA and the agreement between the iDXA and GE Lunar Prodigy densitometers for the measurement of regional- and total-body bone and body composition in normal to obese healthy adults

Comment on results:

Compared the whole-body fat mass by dual-energy X-ray absorptiometry (DXA) to measurements by a 4-component (4-C) model.
iDXA precision was less than 2% of the measured value for all regional- and whole-body bone and body composition measurements with the exception of arm fat mass (2.28%). We found significant differences between iDXA and Prodigy (p < 0.05) whole-body and regional bone, fat mass (FM), and lean mass, with the exception of hip bone mass, area and density, and spine area. Compared to iDXA, Prodigy overestimated FM and underestimated lean mass. However, compared to 4-C, iDXA showed a smaller bias and narrower limits of agreement than Prodigy.