BMD Measurement and Precision: A Comparison of GE Lunar Prodigy and iDXA Densitometers
Objective:
1. To assess (BMD comparability and precision using Lunar Prodigy
and iDXA densitometers
2. to investigate the utility of supine forearm measurement with iDXA
Comment on results:
BMD at the L1L4 spine, total proximal femur, and femoral neck was very highly correlated (r2 ? 0.98) between densitometers, as was the one-third radius site (r2 5 0.96).
BMD precision was similar between instruments at the L1L4 spine, mean total proximal femur, and femoral neck
Finally, one-third radius BMD measurements in the supine vs seated position on the iDXA were highly correlated (r2 = 0.96). In conclusion, there is excellent BMD correlation between iDXA and Prodigy densitometers. Similarly, BMD precision is comparable with these two instruments