Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements

Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements

Objective:

1. Evaluate the agreement of bone mineral measurements between GE Healthcare Lunar Prodigy and iDXA narrow-angle fan beam densitometers;
2. Compare cross-calibration results acquired fromhuman and phantomdata;
3. Calculate potential correction coefficients for Prodigy results to match with those of iDXA;
4. Evaluate the effect of body anthropometry parameters on the agreement of BMD measurements between the instruments.

Comment on results:

The bone mineral density (BMD) measurements with
these two scanners showed a high linear association at all tested sites