Cross-Calibration of Prodigy and Horizon A Densitometers and Precision of the Horizon A Densitometer

Cross-Calibration of Prodigy and Horizon A Densitometers and Precision of the Horizon A Densitometer

Objective:

to enable a reliable transition for clinical study participants and patients from a GE Lunar Prodigy to a Hologic Horizon

Comment on results:

Correlation coefficients for the BMD and body composition measures on the 2 scanners ranged from 0.94 to 0.99 (p<0.001). When compared with values on the Prodigy, mean BMD on the Horizon A was lower at each skeletal site (0.136 g/cm2 lower at the femoral neck and 0.169 g/cm2 lower at the lumbar spine (L1-4)), fat mass was 0.47 kg lower, and lean mass was 4.50 kg higher. Precision of the Horizon A scans was 1.60% for total hip, 1.94% for femoral neck, and 1.25% for spine (L1-4) BMD. Precision of TBS was 1.67%. Precision of total body fat mass was 2.16%, total body lean mass was 1.26%, appendicular lean mass was 1.97%, and total body bone mass was 1.12%.