Diagnostic performance of vertebral fracture assessment by the lunar iDXA scanner compared to conventional radiography

Diagnostic performance of vertebral fracture assessment by the lunar iDXA scanner compared to conventional radiography

Objective:

To evaluate the diagnostic performance of vertebral fracture assessment (VFA) using the Lunar iDXA scanner

Comment on results:

Results: There was excellent agreement between the two techniques and high diagnostic performance of VFA both on a per-vertebra basis (k score = 0.984, 95 % CI 0.972-0.996, sensitivity 98.68 %, specificity 99.91 %, PPV 98.25 %, NPV 99.93 %) and on a per-patient basis

Conclusion; This study demonstrates that most vertebrae are evaluable using the iDXA scanner, with improved VFA diagnostic performance even in discriminating mild VFs from vertebral deformities. Therefore, VFA may be appropriate as an alternative to conventional radiography in patients at high risk of VF who are undergoing DXA bone densitometry and in the follow-up of osteoporotic patients on treatment.