DXA Measured Distal Femur Bone Mineral Density in Patients After Total Knee Arthroplasty: Method Development and Reproducibility
Objective:
To define regional BMD
variation throughout the femur, suggest standard dual-energy X-ray absorptiometry (DXA) regions of interest
(ROIs) and evaluate BMD reproducibility at these ROIs
Comment on results:
Results: The 5 and 10% ROIs included prosthetic and/or patella, causing high BMD values. Distal femur BMD was lowest at the 15% ROI and was higher (p < 0.05) at each more proximal ROI to 45%, then plateaued from 45% to 75%. BMD reproducibility at these regions was excellent; coefficient of variation (CV) from ?1% to 3.5%. As periprosthetic fractures generally occur in the distal femur, we propose measuring femur BMD using ROIs placed at 15% and 25%. A 60% region could also be used as a highly cortical site. Conclusion: Existing DXA capabilities allow distal femur BMD measurement with good reproducibility. Further research using standardized ROIs to assess distal femur BMD loss after TKA, and interventions to mitigate this loss, is indicated.