In Vivo Precision of the GE iDXA for the Assessment of Total Body Composition and Fat Distribution in Severely Obese Patients
Objective:
To evaluate the precision of the iDXA for total body composition and fat distribution
measurements in severely obese patients
Comment on results:
Excellent Precision for
– Total Body BMC = 1.08%
– Total Body fat free mass = 0.94%
– Total Body fat = 0.90%
– Total Body lean = 1%
– % of Total Body Fat = 0.79%
– Gynoid fat distribution = 1.44%
– Android fat distribution = 1.64%