In-vivo precision of the GE Lunar iDXA densitometer for the measurement of appendicular and trunk lean and fat mass

In-vivo precision of the GE Lunar iDXA densitometer for the measurement of appendicular and trunk lean and fat mass

Objective:

To determine the in vivo precision of the Lunar iDXA for the measurement of appendicular and trunk lean and fat mass in a heterogeneous group of adults

Comment on results:

There was excellent agreement between consecutive scans for measurements of both fat and lean appendicular and trunk composition (R2=0.99).
Precision for body composition at all regions was less than 2% CV. Precision for bilateral appendicular body composition was less than 4% CV for the arms and less than 2.5% CV for the legs. Our findings indicate that the iDXA is a valuable tool for repeat measurements of regional body composition in adults