In vivo precision of the GE Lunar iDXA densitometer for the measurement of total body composition and fat distribution in adults

In vivo precision of the GE Lunar iDXA densitometer for the measurement of total body composition and fat distribution in adults

Objective:

The Lunar iDXA provided excellent precision for total body composition measurements. Research into the effect of body size on the precision of DXA body fat distribution measurements is required.

Comment on results:

Pecision was excellent for all measurements, particularly for
Total body BMC : RMS = 0.015kg ; CV = 0.6%
Lean tissue mass : RMS = 0.244kg; CV = 0.5%
Total fat mass : CV=0.82%
% fat : CV=0.86%
Gynoid fat : RMS = 0.397kg; CV 0.96%
Android fat : RMS = 0.780kg, CV 2.32%
There was good agreement between consecutive measurements for all measurements (slope (s.e.) 0.993–1.002; allR2¼0.99).
Lunar iDXA provided excellent precision for total body composition measurements. Research into the effect of body size on the precision of DXA body fat distribution measurements is required