In vivo precision of the GE Lunar iDXA densitometer for the measurement of total-body, lumbar spine, and femoral bone mineral density in adults.

In vivo precision of the GE Lunar iDXA densitometer for the measurement of total-body, lumbar spine, and femoral bone mineral density in adults.

Objective:

The Lunar iDXA provided excellent precision for BMD measurements of the total body, lumbar spine, femoral neck, and total hip.

Comment on results:

Total body, RMS = 0.007 g/cm² ; CV = 0.63%
lumbar spine, RMS = 0.004 g/cm² ; CV = 0.41%
Total hip, RMS= 0.007 g/cm²; CV= 0.53%,
Femoral neck, RMS= 0.014 g/cm²; CV= 1.36%