Precision errors, least significant change, and monitoring time interval in pediatric measurements of bone mineral density, body composition, and mechanostat parameters by GE lunar prodigy

Precision errors, least significant change, and monitoring time interval in pediatric measurements of bone mineral density, body composition, and mechanostat parameters by GE lunar prodigy

Objective:

to (1) evaluate precision errors for bone mineral density, bone mass and bone area, body composition, and mechanostat parameters,
(2) assess the relationships between precision errors and anthropometric parameters, and
(3) calculate a “least significant change” and “monitoring time interval” values for DXA measures in children of wide age range (5-18yr) using GE Lunar Prodigy densitometer

Comment on results:

absolute precision error values were different for thin and standard technical modes of DXA measures and depended on age, body weight, and height. In contrast, relative precision error values expressed in percentages were similar for thin and standard modes (except total body bone mineral density [TBBMD]) and were not related to anthropometric variables (except TBBMD).