Precision of GE Lunar iDXA for the measurement of total and regional body composition in nonobese adults.

Precision of GE Lunar iDXA for the measurement of total and regional body composition in nonobese adults.

Objective:

To characterize in vivo precision of total body and regional body composition parameters using the GE Lunar iDXA

To evaluate the difference between expert and automatic region-of-interest (ROI) analysis on body composition precisio

Comment on results:

Total body precision for
BMC: 0.5%
fat mass = 1%
lean mass = 0.5%
Regional body composition precision error was less than 2.5% CV for all regions except arms (arms, legs, trunck, androîd, gynoÎd)
Significant correlation between auto ROI and expert ROI but statistically significant differences were found between auto and manual ROI