Precision of Lunar Dual-energy X-ray Absorptiometry (iDXA) in measuring body composition among colorectal cancer patients and healthy subjects

Precision of Lunar Dual-energy X-ray Absorptiometry (iDXA) in measuring body composition among colorectal cancer patients and healthy subjects

Objective:

to estimate precision of the Dual-energy X-ray absorptiometry (Lunar iDXA, GE Healthcare software enCORE version 16) in measuring body composition in CRC patients and healthy subjects

Comment on results:

Precision error expressed as coef?cient of variation (% CV) of VAT and SAT were estimated to be 3.56% and 3.28% among CRC patients, and 5.30% and 3.46% among healthy subjects.
Estimated precision errors for body masses in the total region ranged between 0.49-1.01% and 0.40-0.88% in CRC patients and healthy subjects, respectively.
Least significant change (LSC) in VAT mass, SAT mass, FM and LM were 140.9 g, 121.4 g, 637.0 g and 701.0 g, respectively, among CRC patients. Among healthy subjects the LSC in VAT, SAT, FM and LM were 80.93 g, 98.90 g, 484.0 g and 618.0 g, respectively.