Precision of the GE Lunar Total Body-Less Head Scan for the Measurement of Three-Compartment Body Composition in Athletes

Precision of the GE Lunar Total Body-Less Head Scan for the Measurement of Three-Compartment Body Composition in Athletes

Objective:

To investigate the precision of a new total-body-less-head (TBLH) DXA scan for three-compartment body composition measurement in athletes, with outcomes compared to the standard total-body DXA scan

Comment on results:

TBLH precision errors [root mean squared-standard deviation, RMS-SD (coefficient of variation, %CV)] were bone mineral content (BMC): 15.6 g (0.5%), lean mass (LM): 254.3 g (0.4%) and fat mass (FM): 199.4 g (1.3%). These outcomes compared favourably to the precision errors derived from the standard total-body scan [BMC: 12.4 g (0.4%), LM: 202.2 g (0.4%), and FM: 160.8 g (1.1%)]. The TBLH scan resulted in lower BMC (-19.5%), LM (-6.6%), and FM (-4.5%) compared to the total-body scan (BMC: 2,308 vs. 2,865 g; LM: 46,954 vs. 50,276 g; FM: 15,183 vs. 15,888 g, all p<0.005)