Precision of total and regional body fat estimates from dual-energy X-ray absorptiometer measurements
Objective:
To evaluate the precision of both total %fat and all the regional %fat measures acquired from both the Lunar Prodigy and Lunar iDXA software.
Comment on results:
Results: The CV for total body %fat was 1.9% and 0.9% for the Prodigy and iDXA, respectively. CV’s for %fat measures at regional sites ranged from 1.2-4.4% for the Prodigy measures and 0.9-2.4% for the iDXA measures. The ICC for both devices ranged from 0.990 to 0.999.
Conclusion: Monitoring the status of body composition changes with age is gaining more clinical acceptance. Thus, it is important that practitioners use measures that are both precise and accurate. The findings from the current study add support that DXA measurements can be used with a high level of confidence for serial testing of patients.