Objective:
to explore the reliability and precision of body compartment of weight-stable adults,
taken on the GE iDXA
Comment on results:
The CV was higher for all measures over three months (range 0.85.9%) compared with same-day precision-scans (all < 2%).
The CV for VAT was considerably higher than all other body compartments (42.2% three months, 16.2% same day scanning). High levels of variability were demonstrated in VAT mass and volume, 11.8% with same day scanning, and greater than 30% over three months. The highest variability over the three month period was observed in participants with a lower volume of VAT.
Data obtained from participants with VAT mass levels below 500 g should be interpreted with caution.
Results obtained from the 51 participants who completed the three-month reliability study indicate that reliable measures of VAT mass can only be obtained in people with >2 kg android fat mass or 500 g VAT mass. As BMI was positively correlated with VAT mass and negatively associated with variability in VAT, it may serve as a useful indicator when planning nutrition and lifestyle studies aimed at reducing VAT.
The GE Lunar CoreScan application estimate can be improved by adding WC