Spine Trabecular Bone Score Precision, a Comparison Between GE Lunar Standard and High-Resolution Densitometers

Spine Trabecular Bone Score Precision, a Comparison Between GE Lunar Standard and High-Resolution Densitometers

Objective:

To evaluate TBS reproducibility and explored differences between Lunar Prodigy and iDXA densitometers.

Comment on results:

Results: 1-L4 BMD and TBS precision was similar on iDXA and Prodigy (BMD coefficient of variation = 1.9% and 1.5% and TBS coefficient of variation = 1.4% and 1.6%, respectively). Prodigy-to-Prodigy TBS values were highly correlated (R2 = 0.85 with bias of ?0.010 TBS units). Agreement was less robust comparing Prodigy with iDXA instruments (TBS R2: 0.72–0.81 with biases of 0.012–0.034 TBS units).

Conclusion: TBS precision is comparable to that of BMD and does not differ between men and women. Additionally, in these cohorts, slight TBS differences were observed between iDXA and Prodigy scans. These data suggest a potential difference between densitometer models perhaps due to higher iDXA image resolution